1. Calibration
Select One: Maximum temperature limits for each ANSI type thermocouple materials are given (see “more info” button), however these will vary depending on sheath materials, application atmospheres, wire gauge size, insulating materials and epoxy composition.

2. Junction
Tip Style: Tip Styles: The application dictates the choice.   e.g. Fast response time = Exposed Tip.

(G)- TC measuring junction is one with the sheath so they have the same electrical potential.

(U) – TC measuring junction is insulated from the sheath to eliminate outside electrical interference.
Type:

3. Element Type
Select One: Up to three separate thermocouples in a single sheath.  Separate ungrounded junctions can be wired to different instruments.

4. Sheath Material
Select One: The application dictates the choice of sheath materials.  304SS and 316SS are most commonly used but temperatures and corrosive conditions can necessitate the use of other materials.

5. Sheath Diameter
Select One: The Mineral Insulated thermocouples come in a number of sizes but the choice is greatly influenced by the maximum temperature of the application.  For high temp applications bigger is better.

6. Sheath Length "L"
Enter Inches: Although the accuracy of the thermocouple is not affected by the length of the sheath consult the factory when the probe length exceeds 120 inches.

7. Strip Length "A"
Enter Inches:  

8. Fittings
Select Type: Fittings of all thread sizes, materials and design are available for most applications.
 
(For Compression Fittings and Fixed Bushing only)
Select NPT:

9. Fitting Location "F"
Enter inches from tip ( "N" for None ) Measured from the first thread to the tip of the sheath.  Allow for the thread-in insertion length of the fitting when determining location.

10. Special Options
N None W Weld Pad
A Special Limits of Error X X-Ray Junction
I SS ID Tag Y Certificate of Conformance
L Mating Connector 2 High Temp Oxidation Coating
O Seperate Ungrounded Junctions 3 Test at Process Temp
P Electro-etching 5 Soak Test
R Faster Response Construction 7 Rigid Tube Construction
T Coated Probe 8 No Heat Shrink
    Z Other

 


11. Bend Angle
Select One: Mineral insulated thermocouples should not be bent on a radius smaller than twice the outside sheath diameter.

12. Bend Location "C"
Enter inches from tip ( "N" for None ) There are minimum bend locations for all Mineral Insulated sensors – dependent on sheath diameter.
 